How to calibrate zero – Dwyer UXF2 User Manual

Page 48

background image

- 38 -

3.10. How to calibrate zero

Completely close the valves upstream and downstream of the flow meter before calibrating zero.

Notes

1. If there is no valve or if the fluid flow cannot be stopped, select "CLEAR" when "ZERO ADJUST". In this

case, the zero point may slightly be off and manual zero point cannot be set.

2. If parameters are set at "PROTECTION ON", select "PROTECTION OFF" beforehand.
3. SET ZERO: Saves the current flow status as zero. CLEAR: Sets the calibration value for zero point to "0".

The following is the zero point adjustment procedure from measurement mode.

Keying LCD

indication/comment

key pressed twice.

1st line: [OUTPUT SETUP]

ENT

key pressed.

1st line: [ZERO ADJUST]. 2nd line: [CLEAR]. * As selected currently.

ENT

key pressed.

2nd line blinks.

key pressed.

2nd line: [SET ZERO].

ENT

key pressed to register.

On 2nd line about 1 sec, [**COMPLETE**] is displayed, and zero calibration is
performed (Note 4).

ESC

key pressed.

1st line: [OUTPUT SETUP]

key pressed 3 times.

Measurement mode is resumed.

Note: 4. If [CLEAR] has been selected and executed at this time, any saved manual zero calibration

value will be cleared.