beautypg.com

Mounting of small detector on pipe – Dwyer UXF2 User Manual

Page 39

background image

- 29 -

3.6.7. Mounting of small detector on pipe

The small type detector is mounted on pipe with a diameter of ø50 to 300 (V method) or ø300 to 600 (Z method) for
measurements.

3.6.7.1. Mounting of detector (SX3-C0)

Mount the detector using the following procedure.
For mounting, have a ruler, scale, or caliper handy for measurement.

(1) Loosen the retaining knob A (4 places), slide the

detectors to match the mounting dimension, place a
scale on the mounting dimension reference surface C
and adjust the separation distance (dimension). Then
tighten the retaining knobs “A”.

Fig. 3-26

(2) Spread silicone filler over the whole transmitting side

of the detector. Ensure enough silicone is applied
smoothly without bubbles and thoroughly to prevent
air gaps.
Clean the surface of the pipe as required and mount
the detector.

Fig. 3-27

(3) Raise the end of the pipe fitted with the detector, and

attach the yellow ring (on the chain) to the hook.

Fig. 3-28

BULLETIN F-70