How to set the piping parameters – Dwyer UXF2 User Manual
Page 87
- 77 -
4.6.4. How to set the piping parameters
Description
z
Set the parameters of piping and fluid to be measured to determine the sensor unit spacing.
Set the following parameters, calculate the sensor unit spacing value and then, to match the result, install the
sensor units into the frame.
z
Unless the sensor units are spaced accurately, the measurement may demonstrate excessive error.
z
The received signal waveform may be abnormal as a result of incorrect piping or fluid parameters.
Setting item
1. Pipe outer diameter
: 10 to 650 [mm] (factory set at 60.00 [mm]).
2. Pipe material
: PVC (factory set), PVDF, PEEK, PP, CARBON STEEL, STAINLESS STEEL,
COPPER, others (sound velocity 1000 to 3700 m/s).
3. Wall thickness
: 0.1 to 50.00 [mm] (factory set at 4.50 [mm]).
4. Lining material
: NO LINING (factory set), TAR EPOXY, MORTAL, RUBBER, TEFLON,
PYREXGLASS, others (sound velocity: 1000 to 3700 [m/s]).
5. Lining material thickness
: 0.01 to 50.00 [mm].
6. Fluid to be measured
: WATER, SEA WATER, others (sound velocity: 500 to 2500 [m/s])
7. Kinematic viscosity
: 0.0001 to 999.9999
u10
-6
[m
2
/s] (factory set at 1.0038
u10
-6
[m
2
/s])
8. Detector mounting method : V method (factory set), Z method
9. Detector type
: SX3A (factory set), SX3B, SX3C, SX3D
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1)
Typical operation Carry out setting for measuring the flow rate of water flowing through PVC pipe (for tap water)
having 100 mm of nominal diameter.
Keying order
Description
Indication
ź
Pressed 3 times to indicate "MEASURE SETUP".
MEASURE SETUP
ENT
ź
Pressed to indicate "SYSTEM UNIT".
SYSTEM UNIT
METRIC
ź
Pressed 3 times to indicate "PIPE PARAMETER".
PIPE PARAMETER
S= 16 ( 48mm)
ENT
ź
Pressed to indicate "OUTER DIAMETER".
OUTER DIAMETER
60.00 mm
ENT
ź
Pressed to blink the cursor.
OUTER DIAMETER
0
60.00 mm
ź
Operated to compose "114" because, from piping data in
Section 6.6., the outer diameter of polyvinyl chloride pipe
(tap water size) is 114 mm.
OUTER DIAMETER
11
4
.00 mm
Pressed to register the outer diameter.
OUTER DIAMETER
½½ COMPLETE ½½
Ļ
ENT
ź
ź
ź
ź
--- “OUTER DIAMETER” registered.---
OUTER DIAMETER
114.00 mm
BULLETIN F-70