4.5.2. Setting the damping

Description
z

Used for smoothing the variation of measured value.
A time constant is set (response time of about 63%).

Settable range: 0 to 100 sec in 1 sec steps.

Note: In case 0 sec. is set, response time becomes:

• System cycle:

0.2 sec

• Dead time:

less than 0.2 sec

• Time constant:

0.1 sec

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).

Typical operation Change the damping from 5 to 20 sec.

Keying order

Description

Indication

ź

Pressed twice to indicate "OUTPUT SETUP".

OUTPUT SETUP

ENT

ź

Pressed to indicate "ZERO ADJUST".

ZERO ADJUST
SET ZERO

ź

Pressed to indicate "DAMPING".

DAMPING
5 sec

ENT

ź

Pressed to blink the cursor.

DAMPING

0

05 sec

ź

Operated to select "20".

DAMPING
02

0

sec

Pressed to register it.

DAMPING

½½ COMPLETE ½½

Ļ

ENT

ź
ź
ź
ź

--- Damping registered.---

DAMPING
20 sec

ESC

Press

ESC

key once, and

key 3 times to resume

the measurement mode.

0.000 m/s
0.000 m3/h

BULLETIN F-70