How to compensate the measurement value – Dwyer UXF2 User Manual

Page 83

background image

- 73 -

100%

Flow rate

Output

4.5.8. How to compensate the measurement value

Description
z

Used for compensating the measurement value.
The span can be compensated by a preset reading percentage.
Compensatable range: ±200%

The output value (reading, analog output and total output) is
computed by the following expression.

Output =

measurement value

u [compensation value %]

100

For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).

Typical operation Compensate the measurement value by 1 %.

Keying order

Description

Indication

ź

Pressed twice to select the OUTPUT SETUP mode.

OUTPUT SETUP

ENT

ź

Pressed to indicate ZERO ADJUST.

ZERO ADJUST
CLEAR

ź

Pressed 7 times to indicate "CALIBRATION SPAN".

CALIBRATION SPAN
100.0 %

ENT

ź

Pressed to blink the cursor.

CALIBRATION SPAN

1

00.0 %

ź

Pressed twice to move the cursor.

CALIBRATION SPAN
10

0

.0 %

ź

Pressed to select "1".

CALIBRATION SPAN
10

1

.0 %

Pressed to register it.

CALIBRATION SPAN

½½ COMPLETE ½½

Ļ

ENT

ź
ź
ź
ź

--- Compensation value registered.---

CALIBRATION SPAN
101.0 %

ESC

Press

ESC

key once, and

key 3 times to resume

the measurement mode.

0.000 m/s
0.000 m3/h

BULLETIN F-70