beautypg.com

Yokogawa µR20000 User Manual

Page 61

background image

IM 04P02B01-01E

2-6

Dot Model

+/A

–/B

b

+/A

–/B

b

Channel 1
Channel 2
Channel 3
Channel 4
Channel 5
Channel 6

Channel 1

Channel 3

Channel 5

Channel 2

Channel 4

Channel 6

+A

–/B

b

Screw input terminal

Clamped input terminal

(/H2 option)

Channel location on the input terminals

Channel

1 to 6

Channel

7 to 12

Channel

13 to 18 Channel

19 to 24

1.

Turn OFF the recorder and remove the terminal cover.

2.

Connect the signal wires to the terminals.

Clamped input terminal

Screw input terminal

Philips

screwdriver

Flat-blade

screwdriver

Crimp-on lug with

insulation sleeves

Signal wire

Signal wire

Note

Input signal wires of diameter less than or equal to 0.3 mm may not be secured firmly for
clamped terminals (/H2). Fold over the conducting section of the wire, for example, to make
sure that the wire is securely connected to the clamped terminal.

3.

Replace the terminal cover and fasten it with screws.

The proper torque for tightening the screws is 0.6 N•m.

2.3 Input Sgnal Wrng