beautypg.com

4 calibrating the recorder, 4 calibrating the recorder -4, 4 calbratng the recorder – Yokogawa µR20000 User Manual

Page 222: Requred instruments, Calbraton procedure

background image

IM 04P02B01-01E

11-4

11.4 Calbratng the Recorder

Calibrate the measured value against the input.
It is recommended that the recorder be calibrated once a year to assure its measurement
accuracy.
For details regarding calibration, contact your nearest YOKOGAWA dealer.

Requred Instruments

A calibration instrument with appropriate resolution is required for calibrating the recorder.

Recommended Instrument

• DC voltage standard: Model 9100 by FLUKE or equivalent

Man Specfcatons

Output accuracy: ±(0.005% + 1 µV)
• Decade resistance box: Yokogawa Meters & Instruments Model 2793-01 or equivalent

Man Specfcatons

Accuracy of output range 0.1 to 500 Ω: ±(0.01% + 2 mΩ)
Resolution: 0.001 Ω
• 0°C standard temperature device: ZC-114/ZA-10 by Coper Electronics or equivalent

Man Specfcatons

Standard temperature stability accuracy: ±0.05°C

For information on purchasing the calibration instruments, contact your nearest
YOKOGAWA dealer.

Calbraton Procedure

1.

Wire the recorder and the calibration instrument as shown in the following figure,

and adequately warm up the instruments (the warm-up time of the recorder is at
least 30 minutes).

2.

Check that the operating environment such as ambient temperature and humidity

is within the normal operating conditions (see section 12.6).

3.

Apply appropriate input signals corresponding to 0%, 50%, and 100% of the input

range and calculate the errors from the readings.
If the error is outside the accuracy specifications, contact your nearest YOKOGA-
WA dealer.

Note

For thermocouple inputs, you must measure the temperature of the input terminal and apply a
voltage taking into account the reference junction temperature.

DC Voltage Measurement

DC voltage standard

+

Input terminal

L
N

µR20000

+/A

–/B