beautypg.com

Yokogawa µR20000 User Manual

Page 135

background image

IM 04P02B01-01E

6-20

MSG format

Choose a message format. Measured values,* message strings, and date/time can be
set arbitrarily.

47:H:M L16

CHR:Date/Time

CHR:01...

CHR:0A...

CHR:L01...

CHR:Date/Time

:

Measured value

Computed value (/M1 option)

Message string (number of characters)

Date/Time

Comment

CHARACTER key

Example: Based on the message 1 string of the standard function, you can set a

combination with message format 1 and print it out as message 1 including
measured values.*

10th and 11th character of the

message string (9 characters

already used, so this represents

the next characters)

Standard function

Message 1 string

:

Process-1°C

Message 1 printout result

:

06/30 10:10 Process-1 134.8°C

Message format 1

:

M/D H:M L09 01L02

Measured value on CH1 (no units)

Space (number of characters: 1)

Space
1-9th characters of the message string

Date/time

* Computed values can also be included on models with the /M1 option.

• Date/Tme Format and Number of Characters Used

H:M

(Hour:Minute)

5 characters

H:M:S

(Hour:Minute:Second)

8 characters

M/D H:M

(Month/Day Hour:Minute)

11 characters

M/D H:M:S

(Month/Day Hour:Minute:Second)

14 characters

D/M H:M

(Day/Month Hour:Minute)

11 characters

D/M H:M:S

(Day/Month Hour:Minute:Second)

14 characters

D.M H:M

(Day.Month Hour:Minute)

11 characters

D.M H:M:S

(Day.Month Hour:Minute:Second)

14 characters

M.D H:M

(Month.Day Hour:Minute)

12 characters

M.D H:M:S

(Month.Day Hour:Minute:Second)

15 characters

Y/M/D H:M:S

(Year/Month/Day Hour:Minute:Second)

19 characters

M/D/Y H:M:S

(Month/Day/Year Hour:Minute:Second)

19 characters

D/M/Y H:M:S

(Day/Month/Year Hour:Minute:Second)

19 characters

D.M.Y H:M:S

(Day.Month.Year Hour:Minute:Second)

19 characters

M.D.Y H:M:S

(Month.Day.Year Hour:Minute:Second)

20 characters

• Format of Measured Values and Computed Values, and Number of Characters Used

• The format for measured values is 01, 02, ...24 (depending on the model). No units

are added. 7 characters are used.

• The format for computed (/M1 option) values is: 0A, 0B, 0C, 0D, 0E, 0F, 0G, 0J, 0K,

0M, 0N, 0P, 1A, 1B, 1C, 1D, 1E, 1F, 0G, 1J, 1K, 1M, 1N, and 1P (depending on the
model). No units are added. 9 characters are used.

• Character Strng Format

The format for message strings is L01 (1 character), L02 (2 characters), .... L16 (16

characters).

In the example above, L09 specifies “Process-1” and L02 specifies “°C .”

Enabling the message format: Section 7.24

6.15 Settng the Message Format (/BT1 Opton)