beautypg.com

Explanaton – Yokogawa µR20000 User Manual

Page 129

background image

IM 04P02B01-01E

6-14

Explanaton

Settng Correcton Ponts (Measured Values) and Correcton Values

• Using revise values

Calibration point

B4

A3

A2

A1

A4

B3

B2

B1

Scale value

Correction

value

Measured

value

Correction using revise values

Correction point (measured value) = A

Correction value = B – A

• Using absolute values

Calibration point

B4

A3

A2

A1

A4

B3

B2

B1

Scale value

Correction

value

Measured

value

Correction using absolute values

Correction point (measured value) = A

Correction value = B

Example of a two-pont correcton

Calbraton

Pont

Corrected

Value

Measured

value (A)

Correcton Value

Revse Value

(B – A)

Absolute

Value (B)

1

10.0°C

9.8°C

0.2°C

10.0°C

2

90.0°C

90.5°C

– 0.5°C

90.0°C

Setup Condtons of Measured and Correcton Values

• Setting range of measured and correction values when using revise values
A1 < A2 ≤ A3 ≤ ... ≤ A16
(A1 + B1) < (A2 + B2) < (A3 + B3) < ... < (A16 + B16)
A and B are within the measurable range or scaling range of the range type*

* Within the scaling range: –5% to 105% of the scale value or –20000 to 30000 (the decimal

place is the same as the setting for the scale value)

• Setting range of measured and correction values when using absolute values
A1 < A2 ≤ A3 ≤ ... ≤ A16
B1 < B2 < B3 < ... < B16
A and B are within the measurable range or scaling range of the range type*

* Within the scaling range: –5% to 105% of the scale value or –20000 to 30000 (the decimal

place is the same as the setting for the scale value)

Note

• Make sure that the leftmost value is less than the rightmost value for span and scale.
• Calibration correction cannot be set on ON/OFF input (DI), delta computation, and square

root computation channels. In addition, calibration correction cannot be used on all channels
if the bias function is enabled.

Enabling the calibration correction function: Section 7.15

Setting the calibration correction function: Section 7.23

6.13 Performng Calbraton Correcton (/CC1 Opton)