6-12

6.12 Applyng a Bas on the Measurng Input Sgnal

This section explains the details of setting a bias on the measuring input signal.
If you change the input range, set the bias again.

Procedure

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Press the

key or

SHIFT +

key to select

Bas and then press the key.

3.

Set each item and press the key.

Use the

key or

SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.
If you press the

ESC key in the middle of the operation, the settings entered up to

then are cancelled, and the display returns to a higher level menu.

4.

When the

Settng complete screen appears, do either of the following:

Press the key to set other channels.
If you are done, press the

ESC key.

5.

Hold down the

MENU

key for 3 seconds to return to Operation mode.

Select whether to use the bias.

Set the bias value.

Set the channel range.

The new setting takes effect.

First channel

Last channel

ESC/?

(When bias is On)

Set=Bias

CH=01-01

Bias=Off

Bias= 0.000

01-01 Channel

Setting complete

Explanaton

Channel Range

The only case when a bias can be set on multiple channels simultaneously is when the
channels are set to the same range type (for example channel 01 and 02 are set to 2 V
range). For channels on which scaling is set, the channels must be set to the same range
type, same span values, and same scaling values.

Bas

The range of bias that can be specified is ±10% of the span of the measurable range at
the specified range type. For example, the range is –0.4 V to 0.4 V for the range type 2 V.
For channels on which scaling is set, the range is ±10% of the scaling span.

Note

A bias cannot be set on channels set to ON/OFF input (DI).

Enabling the bias function: Section 7.15