Software (sold separately) – Yokogawa µR20000 User Manual
Page 3
IM 04P02B01-01E
Recorder’s Verson and Functons Descrbed n Ths
Manual
The contents of this manual corresponds to the recorder with version 1.31.
µR20000 Versons and Functons
Verson
Suffx Code Added or Modfed Functons
Reference
1.11 or earlier
–
–
–
1.21
-2
(Added) Language support (German and French).
Section 1.9
/CC1
(Added) Calibration Correction.
Section 1.2
/H5x
(Added) Portable type.
Sections 2.5 and 12.7
/P1
(Added) 24 VDC/AC power supply operation.
Section 2.5
1.31
–
(Added) Customized menu.
Section 1.9
–
(Added) Modbus register (40301 to 40348).
Communication manual
/BT1
(Added) Header printout.
Section 1.4
• Checking the Version Number
You can check the version number on the System display.
The System display cannot be shown at the factory default condition.
First, register the System display to the display screen.
• Procedure of registering the System display to the display screen: See section 8.2.
• Procedure of displaying the System display: The screen switches each time the
DISP
key is pressed. Press
the
DISP
key repeatedly until System display is shown. The displayed contents on the System display
switches every 3 seconds. Check the number shown by the “Version:” item.
Software (Sold Separately)
The table below shows the relationship between the RXA10 Configuration Software revisions and the µR20000
recorder versions.
Recorder verson
1.11 or earlier
1.21
1.31
RXA10 Confguraton
Software revson
R2.01
Yes
Yes
Limited
R3.01
Yes
Yes
Yes
Yes:
Compatible
Limited: The new functions of the recorder cannot be configured from the RXA10.
Note
Set the recorder version in the RXA10 Configuration Software to display the setup items that