beautypg.com

Yokogawa µR20000 User Manual

Page 209

background image

IM 04P02B01-01E

9-26

Explanaton

Mode

Sets the type of report data to be printed.
AVE: Prints the average value over the interval.
MIX: Prints the minimum, maximum, and average values over the interval.
SUM: Prints the sum value over the interval. If SUM is selected, set the sum scale.
MIN: Prints the minimum value over the interval.
MAX: Prints the maximum value over the interval.
INST: Prints the instantaneous value.

SUM scale

Set this item only when SUM is selected.
For the definition of the term “sum scale,” see page 7-14.
Off: Simply sums the computed values.
/s: Sums by converting the computed values to a value over 1 second.
/min: Sums by converting the computed values to a value over 1 minute.
/h: Sums by converting the computed values to a value over 1 hour.
/day: Sums by converting the computed values to a value over 1 day.

Setting the periodic printout interval and the type of data to be

printed: Section 7.8

9.15 Changng the Type of Report Data Prnted n Perodc Prntout