7.15 Enablng the Bas, Low-Cut, Alarm Delay,
Calbraton Correcton (/CC1 Opton) Functons
This section explains the details of enabling/disabling the bias, low-cut, alarm delay, and
calibration correction (/CC1 option) functions.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changng the Settngs
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or
SHIFT +
key to select
Personalze and then press the
key.
4.
Press the key with
Add functon shown on the screen.
5.
Set each item and press the key.
Use the
key or
SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the
ESC key in the middle of the operation, the settings entered up to
then are cancelled, and the display returns to a higher level menu.
6.
When the
Settng complete screen appears, do either of the following:
To correct the setting, press the key.
If you are done, press the
ESC key.
Enable/Disable the bias function.
Enable/Disable the low-cut function on
channels set to square root computation.
Enable/Disable the low-cut function on
channels set to 1-5V.
The new setting takes effect.
Enable/Disable the calibration correction
function.
(/CC1 option)
ESC/?
Personalize
Setting complete
Basic=Personalize
Mode=Add function
Bias=Not
SQRT low-cut=Use
1-5V low-cut=Not
Calibration=Use
Enable/Disable the alarm delay function.
Alarm delay=Not