Yokogawa µR20000 User Manual
Page 227
11-9
IM 04P02B01-01E
Ma
ntenance
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
7.
Press the
ESC key to finish the adjustment. The screen returns to the
Basc=P_Adj screen.
Select Hysteresis, Zero, or Full.
Change the value to adjust the pen
position.
The dot printing position moves.
ESC/?
Basic=P_Adj
P_Adj=Hysteresis
P_Adj
Setting complete
Hysteresis=-2
Applyng the Changes and Returnng to Operaton Mode
Press the
key or
SHIFT +
key to select
End and then press the key. Press
the
key to select
Store and then press the key. The changes are applied, and
the screen returns to Operation mode. If you select
Abort and press the key, the
changes are discarded, and the screen returns to Operation mode. Press the
ESC key
to return to the
Basc= screen.
Note
When the Customized Menu is enabled, the operation is different.
• When “P_Adj” n the Customzed Menu s On
1.
Perform steps 1-4 of “Adjusting the Dot Printing Position.”
2.
When the instrument enters Basic Setting mode, “Menu kind” appears. Select “Free”
then press a
key.
3.
Perform the steps starting with step 5 of “Adjusting the Dot Printing Position.”
• When “P_Adj” n the Customzed Menu s Off
See “Disabling the Customized Menu” or “Changing the Settings without Disabling the Basic
Setting Mode Lock.”
Changing the settings without releasing the basic setting mode lock:
Section 7.22
11.6 Adjustng the Dot Prntng Poston (Dot Model)