5v input – Yokogawa µR20000 User Manual
Page 107
IM 04P02B01-01E
5-6
1-5V Input
Select 1-5V.
Set the channel range.
Set the left span value.
Set the right span value.
Set the decimal position and left scale
value.
Set the right scale value.
Turn On or Off the low-cut function.
The new setting takes effect.
(Select the right scale value.)
ESC/?
Set=Range
CH=01-01
Mode=1-5V
Span left= 1,000
Span right= 5,000
Scale left= 0.00
Scale right= 200.00
Low-cut=Off
01-01 Channel
Setting complete
First channel
Last channel
Descrpton
• Span Left and Span Rght
Selectable span left range: 0.8 to 1.2 V
Selectable span right range: 4.8 to 5.2 V
• Scale Left, Scale Rght, and Decmal Poston
Selectable range (mantissa): –20000 to 30000
Example: The value in the range of –100.00 to 350.00 cannot be specified. The
mantissa of scale right is 35000, which exceeds the upper limit of 30000.
Set in the range of –100.0 to 350.0.
Decimal position: XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
The decimal position is the position specified by scale left.
Example: Change 0.00 to 0.0.
Press the
key to move the cursor to the last digit of 0.00.
Space
Press the key to change the value to 0.0 .
Make sure that scale left is less than scale right.
• Low-cut
On: Cuts values below 0% of the recording span (set to 0% (scale left value)).
Off: Does not apply the low-cut function.
Enabling the low-cut function: Section 7.15
5.1 Settng the Input Range