1.6 Computaton Functon (/M1 Opton)

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

• Communcaton Input Data (Pen model: C01 to C08, dot model: C01 to

C24)

This data type is set using the communication interface (/C7 or /C3 option).
Range of values (maximum significant digits is 5):
–9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
For the procedure of setting the communication input data, see the µR10000/µR20000

Communication Interface User’s Manual (IM04P01B01-17E).

• Status of the Remote Control Input Termnal (D01 to D05, /R1 opton)

The status of the remote control input signal (1 or 0) can be used in the computation.

Use D01 to D05 (the number following the letter D is the remote control input terminal
number) to specify the status in the equations.

Setting the computing equation, constant, and unit: Sections 9.2 to 9.4

Setting the TLOG timer: Sections 9.6 and 9.12

Handng of the Unt n Computatons

In computations, measured/computed values are handled as values without units. For
example, if the measured value of channel 01 is 20 mV and the measured value of
channel 02 is 20 V, the computed result of 01 + 02 is 40.

Recordng Computaton Channels

Trend Recordng

Trend recording is possible by setting the recording span for each computation channel.
The recording span of computation channels can be set in the range of –9999999 to
99999999 excluding the decimal.
Zone recording and partial expanded recording are also possible.

Assgnment of Computaton Channels to the Pens (Pen Model)

Trend recording is possible by assigning computation channels to arbitrary pens. Only
display or printout is possible for measurement channels or computation channels that
are not assigned to a pen. Assignment of pens is possible only on models with the
computation function.

Pen

1 2 3 4

Channel 01 02 03 04 0A 0B 0C 0D 0E 0F . . . . 1P

Measurement channels Computation channels

Changing the channel assignment of recording pens: Section 9.14

Changng the Channel Recordng Color (Dot Model)

The recording colors of computation channels can be changed. The default recording
colors of channels are indicated below.

Recordng color

Purple

Red

Green

Blue

Brown

Black

Computation channels

0A

0B

0C

0D

0E

0F

0G

0J

0K

0M

0N

0P

1A

1B

1C

1D

1E

1F

1G

1J

1K

1M

1N

1P