Other specfcatons – Yokogawa µR20000 User Manual
Page 250
IM 04P02B01-01E
12-22
Item
Specfcatons
Ambient temperature
With temperature variation of 10°C:
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:
Measurement fluctuation + ±0.2% of the recording span or less
* Excluding the error of reference junction compensation
Power supply fluctuation
With variation within 90 to 132 V and 180 to 264 VAC (50/60 Hz):
Measurement: ±1 digit or less
Recording:
±0.1% of the recording span or less
With variation of ±2 Hz from rated power frequency (power supply voltage 90 to 132, 180 to 264 VAC):
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:
Same as the measurement fluctuation
Magnetic field
AC (50/60 Hz) and DC 400 A/m fields:
Measurement: ±(0.1% of rdg 10 digit) or less
Recording:
±0.5% of the recording span or less
Input source resistance
DC voltage range
With variation of +1 kΩ:
Ranges of 200 mV or less:
Within ±10 µV
Ranges of 2V or more:
±0.1% of rdg
TC range
With variation of +1 kΩ: ±10 µV or less
RTD range
With variation of 10 Ω per wire (resistance of all three wires must be equal):
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:
Measurement fluctuation + ±0.1% of the recording span or less
With maximum difference of 40 mΩ between wires: Approx. 0.1°C (for Pt100)
Mounting position
With a backward tilt of 30° or less
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:
Indication fluctuation + ±0.1% of the recording span or less
Vibration
When a sinusoidal vibration along all three axis at a frequency between 10 to 60 Hz and an
acceleration of 0.2 m/s
2
is applied for 2 hours
Measurement: ±(0.1% of rdg + 1 digit) or less
Recording:
Indication fluctuation + ±0.1% of the recording span or less
Other Specfcatons
Item
Specfcatons
Clock
With a calendar function
Accuracy of clock
±100 ppm, excluding a delay (of 1 second, maximum) caused each time the power is turned on.
DST
The date/time for switching between standard time and DST can be specified.
Memory backup
A built-in lithium battery backs up the setup parameters and runs the clock
(battery life: approximately ten years at room temperature).
Key lock function
Password for releasing the key lock can be set.
Key lock items
RCD key, MENU key, DISP MENU key, FEED key, and function under the FUNC key (Alarm ACK,
computation start/stop, print, message, printout buffer clear, periodic printout reset, and pen or
ribbon cassette exchange)
Internal light
Lights up the recording area of the chart paper. Uses a dedicated LED.
Noise
Machine Noise Information Ordinance 3.GSGV, Jan 18, 1991
Maximum noise level: 60 dB (A) or less (complies with ISO7779)
12.6 General Specfcatons