6.1 Settng the Trend Recordng Interval (Dot

Model)

This section explains the details of setting the trend recording interval (dot printing
interval) on the dot model.

Procedure

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Press the

key or

SHIFT +

key to select

Aux and then press the key.

3.

Press the

key or

SHIFT +

key to select

Trend and then press the key.

4.

Set the dot printing interval and press the key.

Use the

key or

SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.
If you press the

ESC key in the middle of the operation, the settings entered up to

then are cancelled, and the display returns to a higher level menu.

5.

When the

Settng complete screen appears, do either of the following:

To correct the setting, press the key.
If you are done, press the

ESC key.

6.

Hold down the

MENU

key for 3 seconds to return to Operation mode.

Select the trend recording interval

mode.

The new setting takes effect.

ESC/?

Set=Aux

Aux=Trend

Mode=Auto

Trend Interval

Setting complete

Explanaton

Mode

Auto: The trend recording interval is set according to the chart speed in the range of

10 s to 90 s so that the dots do not overlap many times.. The interval is displayed
on the System display (see page 12-9).

Fix: The trend recording interval varies depending on the number of channels

being recorded (see the table below). On models with the computation function
(/M1 option), the trend recording interval varies depending on the number of
measurement and computation channels to be trend recorded.

Number of channels to be recorded Trend recordng nterval

6 or less

10 s

7 to 12

15 s

13 to 18

20 s

19 to 24

30 s

25 to 48

60 s

When set to Auto

When set to Fix

Dot

The distance between dots printed on the chart

paper varies depending on the chart speed.

Chapter 6

Setup Operatons for Convenent Functons (Settng Mode)