Delta computaton – Yokogawa µR20000 User Manual

Page 105

background image

IM 04P02B01-01E

5-4

Delta Computaton

Select Delta.

Set the channel range.

Set the reference channel.

Set the left span value.

Set the right span value.

The new setting takes effect.

ESC/?

Set=Range

CH=02-02

Mode=Delta

Ref.CH=01

Span left= -2.000

Span right= 2.000

02-02 Channel

Setting complete

First channel

Last channel

Descrpton

• Reference Channel

The reference channel must be a channel that is smaller in channel number than the

channel being set.

• Span Left and Span Rght

The range type is set to the same type as the reference channel. The span left and

span right can be set in the range shown below. Span left and span right cannot be
set to the same value.

Input Type

Range Type

Selectable Span Range

DC voltage

20 mV

–20.00 to 20.00 mV

(Volt)

60 mV

–60.00 to 60.00 mV

200 mV

–200.0 to 200.0 mV

2 V

–2.000 to 2.000 V

6 V

–6.000 to 6.000 V

20 V

–20.00 to 20.00 V

50 V

–50.00 to 50.00 V

Thermocouple R

–1760.0 to 1760.0°C

–3168 to 3168°F

(TC)

S

–1760.0 to 1760.0°C

–3168 to 3168°F

B

–1820.0 to 1820.0°C

–3276 to 3276°F

K

–1570.0 to 1570.0°C

–2826 to 2826°F

E

–1000.0 to 1000.0°C

–1800 to 1800°F

J

–1300.0 to 1300.0°C

–1999.9 to 2340.0°F

T

–600.0 to 600.0°C

–1080.0 to 1080.0°F

N

–1300.0 to 1300.0°C

–2340 to 2340°F

W

–1999.9 to 2315.0°C

–4167 to 4167°F

L

–1100.0 to 1100.0°C

–1980 to 1980°F

U

–600.0 to 600.0°C

–1080.0 to 1080.0°F

WRe

–1999.9 to 2400.0°C

–4320 to 4320°F

RTD

PT (Pt100)

–800.0 to 800.0°C

–1440.0 to 1440.0°F

JPT (JPt100) –750.0 to 750.0°C

–1350.0 to 1350.0°F

5.1 Settng the Input Range