IM 04P02B01-01E

7-4

7.2 Changng the Integraton Tme of the A/D

Converter

This section explains the details of setting the integration time of the A/D converter.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).

Procedure

Changng the Settngs

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Hold down the

and

keys simultaneously for 3 seconds to enter Basic

Setting mode.

3.

Press the

key or

SHIFT +

key to select

A/D and then press the key.

4.

Set each item and press the key.

Use the

key or

SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.
If you press the

ESC key in the middle of the operation, the settings entered up to

then are cancelled, and the display returns to a higher level menu.

5.

When the

Settng complete screen appears, do either of the following:

To correct the setting, press the key.
If you are done, press the

ESC key.

Select the integration time of the

A/D converter.

The new setting takes effect.

ESC/?

A/D Integrate

Setting complete

Basic=A/D

Integrate=Auto

Applyng the Changes and Returnng to Operaton Mode

Press the

key or

SHIFT +

key to select

End and then press the key. Press

the

key to select

Store and then press the key. The changes are applied, and

the screen returns to Operation mode. If you select

Abort and press the key, the

changes are discarded, and the screen returns to Operation mode. Press the

ESC key to

return to the

Basc= screen.

Explanaton

Integrate

Selects the integration time of the A/D converter.

Settng

Integraton Tme

Scan Interval

Pen Model

Dot Model

6-dot

12-dot

18, 24-dot

50 Hz

20 ms

125 ms

1 s

2.5 s

2.5 s

60 Hz

16.7 ms

125 ms

1 s

2.5 s

2.5 s

Auto

20 ms or 16.7ms

2, 3

125 ms

1 s

2.5 s

2.5 s

100 ms

1

100 ms

2.5 s

5 s

10 s

1: Selectable only on the dot model.
2: Automatically switch by detecting the power supply frequency.
3: The integration time is fixed to 20 ms if “Auto” is selected when using the DC power supply on a

model with the /P1 option.