Yokogawa µR20000 User Manual
Page 108
5-7
IM 04P02B01-01E
Frequently
Used
Setup
Operat
ons
(Sett
ng
Mode)
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Note
• The displayable/printable range of scale values is –19999 to 30000 excluding the decimal.
• If the leftmost value of the scale is set to –20000 (excluding the decimal), the value when
the low-cut function is applied is displayed as “–Over.”
Square Root Computaton
Select SQRT.
Set the channel range.
Select the range type of DC voltage.
20 mV, 60 mV, 200 mV, 2 V, 6 V, 20 V, or 50 V
Set the left span value.
Set the right span value.
Set the decimal position and
left scale value.
Set the right scale value.
Turn On or Off the low-cut function.
(When low-cut is set to Use)
(When low-cut is On)
Set the low-cut point.
0.0 to 5.0% of the recording span, 0.1 steps.
The new setting takes effect.
ESC/?
Set=Range
CH=01-01
Mode=SQRT
Range=2V
Span left= -2.000
Span right= 2.000
Scale left= 0.00
Scale right= 200.00
Low-cut=Off
Low-cut point=0.0
01-01 Channel
Setting complete
First channel
Last channel
Descrpton
• Input Type n Whch Square Root Computaton Can Be Performed
Square root computation can be performed when set to DC voltage input.
• Span Left and Span Rght
Same as the selectable span range of the DC voltage input.
Make sure that span left is less than span right.
5.1 Settng the Input Range