For the procedure to set the functions, see section 1.10, “Function Setup Guide.”

1-12

1.3 Alarms

Non-Hold/Hold Operaton of Alarm Output Relays

The alarm output relay can be set to operate in the following fashion when the alarm
condition is no longer met.
• Turn off the relay output (non-hold).
• Hold the relay output until the alarm ACK operation is executed (hold).
The setting applies to all alarm output relays.
The default setting is non-hold.

Alarm

Alarm occurrence

Alarm release

Alarm output

relay

Alarm

ACK

Alarm

ACK

Alarm

ACK

or

or

Hold

Non-hold

Activated

Released

Note

• If the reflash alarm is enabled, I01 to I03 are fixed to non-hold operation. Specifying Hold

produces no effect.

• If diagnosis output is enabled, I01 is fixed to non-hold operation. Specifying Hold produces

no effect.

Setting the non-hold/hold operation of alarm output relays: Section 7.1

Alarm ACK Operaton

The alarm acknowledge (alarm ACK) operation releases all alarm indications and relay
outputs (/A1, /A2, /A3, /A4, and /A5 options) that are activated when the alarm indication
or alarm output relay is set to hold operation. This operation can be executed from the
front panel key.

Alarm ACK operation: Section 3.12