App index, Alarm functons, Recordng functons – Yokogawa µR20000 User Manual

Page 51

background image

1-35

IM 04P02B01-01E

Funct

onal
Explanat

on

and
Setup

Gu
de

1.10 Functon Setup Gude

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

Item

Descrpton

Reference

Secton

Moving average (dot model)

Use

Movng_AVE in Basic Setting mode and select Use or Not.

7.11

If Use is selected, set the number of samples of moving average using

Aux >

Movng_AVE in Setting mode.

6.3

If Not is selected, the

Aux > Movng_AVE item does not appear.

Integration time of the A/D converter

7.2

Use

A/D in Basic Setting mode to set the integration time of the A/D converter.

Temperature Unit

Select the temperature unit using

Temperature in Basic Setting mode.

7.25

Alarm functons

Item

Descrpton

Reference

Secton

Alarms for each channel

Use

Alarm in Setting mode.

5.2

Delay high limit alarm/delay low limit alarm

• Alarm type

7.15

Use

Personalze > Add functon > Alarm delay in Basic Mode and select Use or

Not.

If Use is selected, delay high limit alarm (T) and delay low limit alarm (t) become

selectable under

Alarm > Type in Setting mode.

5.2

If Not is selected, alarm delay cannot be selected.
• Alarm delay time

6.10

Use

Aux > Alarm delay tme in Setting mode to set the alarm delay time.

High limit on rate-of-change alarm/low limit on rate-of-change alarm

Use

Alarm in Setting mode and set the channel to rate-of-change alarm.

5.2

Use

Alarm > Increase in Basic Setting mode to set the interval of the high limit on

rate-of-change alarm; use

Alarm > Decrease in Basic Setting mode to set the interval

of the low limit on rate-of-change alarm.

7.1

Use the alarm ACK operation

• Relay operation and display operation

7.1

Set

Alarm > Behavor in Basic Setting mode to Hold and set Alarm > Indcator to

Hold.

• Executing the alarm ACK operation

3.12

Use

Func > Alarm ACK in Operation mode to execute the release of the alarm

output.

Set a hysteresis on the alarm occurrence/release value of high limit alarm and low limit alarm

Use

Alarm > Hysteress in Basic Setting mode to set the hysteresis to be applied to

the high limit alarm and low limit alarm of measurement channels.

7.1

Likewise, use

Alarm > Math Hysteress to set the hysteresis to be applied to the high

limit alarm and low limit alarm of computation channels (/M1 option).

Change the alarm output relay operation

If

Alarm > Act in Basic Setting mode set to De_energize, alarm output relay is

energized during normal operation and de-energized when an alarm occurs.

7.1

Diagnosis output

Use

Alarm > Dagnoss in Basic Setting mode to set the function.

7.1

Recordng functons

Item

Descrpton

Reference

Secton

Chart speed

Use

Chart in Setting mode to set the chart speed.

5.4

Turn trend recording ON/OFF (dot model)

Use

Aux > Prnt out in Setting mode to set whether to trend record on each

measurement channel.

6.6

Change the recording color of measurement channels (dot model).

Use

Color in Basic Setting mode to set the recording color of measurement channels. 7.5

Record by compensating for the pen offset along the time axis (pen model)

Use

Aux > POC in Setting mode to turn On/Off offset compensation.

7.6

Record by setting recording zone

Use

Aux > Zone in Setting mode to set the recording zone for each channel.

6.4

Partial expanded recording Use

Partal in Basic Setting mode and select Use or Not.

7.13

If Use is selected, set the display mode using

Aux > Partal in Setting mode.

6.5

If Not is selected, the

Aux > Partal item does not appear.