Isolaton

Item

Specfcatons

Insulation resistance

Each terminal to ground terminal: 20 MΩ or more (at 500 VDC)

Dielectric strength

Power supply to ground terminal: 1500 VAC (50/60 Hz), 1 minute
Contact output terminal to ground terminal: 1500 VAC (50/60 Hz), 1 minute
Measuring input terminal to ground terminal: 1000 VAC (50/60 Hz), 1 minute
Between measuring input terminals: 1000 VAC (50/60 Hz), 1 minute (except for RTD input terminal)
Remote input terminal to ground terminal: 500 VDC, 1 minute

Grounding

Grounding resistance: 100 Ω or less

Transport and Storage Condtons

Item

Specfcatons

Ambient temperature

–25 to 60°C

Ambient humidity

5 to 95%RH (no condensation)

Vibration

10 to 60 Hz, 4.9 m/s

2

maximum

Shock

392 m/s

2

maximum (in packaged condition)

Supported Standards

Item

Specfcatons

CSA

Certified by CSA22.2 No. 61010-1 (certified by NRTL/C), installation category II

1

, measurement

category II

2

, pollution degree 2

3

* “US” (USA) and “C” (Canada) are indicated to the right and left of the CSA mark, respectively,

to show that NRTL is included.

CE

EMC

EN61326(Emission: Class A, Immunity: Annex A)
EN61000-3-2
EN61000-3-3
EN55011 compliance Class A Group 1

Low voltage directive EN61010-1 compliance, Installation category II, Measurement category II, Pollution degree 2

C-Tick

AS/NZS CISPR 11 compliance Class A Group 1

1 Installation category (overvoltage category) describes a number which defines a transient overvoltage condition. It implies

the regulation for impulse withstand voltage. “II” applies to electrical equipment which is supplied from the fixed installation
like a distribution board.

2 Applies to measuring circuits connected to low voltage installation, and electrical instruments supplied with power from fixed

equipment such as electric switchboards.

3 “Pollution degree” describes the degree to which a solid, liquid, or gas which deteriorates dielectric strength or surface

resistivity is adhering. “2” applies to normal indoor atmosphere. Normally, only non-conductive pollution occurs.