Advanced register: trigger counter, Table 188: advanced register: trigger counter – ALLIED Vision Technologies Pike F-1600 User Manual
Page 367

Configuration of the camera
PIKE Technical Manual V5.1.2
367
Advanced register: trigger counter
The trigger counter feature is controlled by the following advanced feature reg-
ister:
Having this feature enabled, the current trigger counter value (external trigger
seen by hardware) will be inserted as a 32-bit integer value into any captured
image.
Setting the Reset flag to 1 resets the trigger counter to 0: the Reset flag is self-
cleared.
The ON_OFF and LinePos fields are simply mirrors of the time stamp feature.
Settings of these fields are applied to all image stamp features.
Additionally there is a register for direct read out of the trigger counter value.
Register
Name
Field
Bit
Description
0xF1000620 TRIGGER_COUNTER
Presence_Inq
[0]
Indicates presence of this feature
(read only)
Reset
[1]
Reset trigger counter
---
[2..31]
Reserved
0xF1000624 TRGCNT
TriggerCounter
[0..31]
Trigger counter
Table 188: Advanced register: Trigger counter
Note
The 4 bytes of the trigger counter value will be inserted as the
9th to 12th byte of a line.