ND 287

95

II – 3 S

w

itc

h

ing inputs and outputs

Switching outputs at D-sub connection X41

Output signals

Signal level of outputs

Danger to internal components!

The power supply of external circuits must comply with
EN 50178 requirements for low voltage electrical
separation

.

Connect inductive loads only with a quenching diode
parallel to the inductance

.

Danger to internal components!

Use only shielded cables and connect the shield to the
connector housing

.

Pin

Function

14

Display value is zero

15

Measured value is greater than or equal to trigger limit A1

16

Measured value is greater than or equal to trigger limit A2

17

Measured value is less than the lower sorting limit

18

Measured value is greater than the upper sorting limit

19

Error (see "Error messages" on page 60)

Fig. II.60 Open-collector outputs

Signal

Value

Open-collector outputs

Active Low

Delay until signal output

t

V

20 ms

Signal duration of zero

crossover, trigger limit A1, A2

t

0

180 ms

Status

Level

High

U

+32 V; I

10 µA

Low

U

+0.4 V; I

100 mA