HEIDENHAIN ND 287 User Manual User Manual
Page 125

ND 287
125
II – 7 Input and output of par
a
met
e
r list and er
ro
r
compensation table
P52
DATA BIT =
0
Data bits: 0 = 7 bits , 1 = 8 bits
P53
STOP BIT =
0
Stop bit: 0 = 2 stop bits, 1 = 1 stop bit
P54
PARITY BIT =
1
Parity bit: 0 = none, 1 = even, 2 = odd
P55
BLANK LINE =
1
Blank lines: 1 (0 - 99)
P56 DISP.FREEZE =
0
Display freeze: 0 = current, 1 = frozen, 2 = stopped
P60
PRESET = +
0.0000
Value for preset via external signal: 0.0000
P61
A1 ON/OFF =
1
Switching output A1: 0 = off, 1 = on
P62
A2 ON/OFF =
1
Switching output A2: 0 = off, 1 = on
P63
LIMIT A1 = +
0.0000
Value for switching output A1: 0.0000
P64
LIMIT A2 = +
0.0000
Value for switching output A2: 0.0000
P66
PART NO. =
0
Part number selected: 0 is the default setting
P67
LOW.LIMIT 0 =
+
50.0000
LOW.LIMIT 1 =
+
0.0000
LOW.LIMIT 2 =
–
0.0170
LOW.LIMIT 3 =
–
25.0000
LOW.LIMIT 4 =
–
5.0000
LOW.LIMIT 5 =
–
6.0000
LOW.LIMIT 6 =
–
7.0000
LOW.LIMIT 7 =
–
7.0000
LOW.LIMIT 8 =
–
9.0000
LOW.LIMIT 9 =
–
254.0000
Lower tolerance limit for part 0 – part 9: 0.0000 is the default setting
P68
NOM.VALUE 0 =
+
55.0000
NOM.VALUE 1 =
+
0.0000
NOM.VALUE 2 =
+
0.0000
NOM.VALUE 3 =
+
0.0000
NOM.VALUE 4 =
+
1.0000
NOM.VALUE 5 =
+
0.0000
NOM.VALUE 6 =
+
0.0000
NOM.VALUE 7 =
+
0.0000
NOM.VALUE 8 =
+
0.1000
NOM.VALUE 9 =
+
13.9700
Nominal dimension for part 0 – part 9: 0.0000 is the default setting
P69
UP. LIMIT 0 =
+
60.0000
UP. LIMIT 1 =
+
0.0000
UP. LIMIT 2 =
+
0.0170
UP. LIMIT 3 =
+
25.0000
UP. LIMIT 4 =
+
5.0000
UP. LIMIT 5 =
+
6.0000
UP. LIMIT 6 =
+
7.0000
UP. LIMIT 7 =
+
7.0000
UP. LIMIT 8 =
+
9.0000
UP. LIMIT 9 =
+
254.0000
Upper tolerance limit for part 0 – part 9: 0.0000 is the default setting
Parameter
Meaning