HEIDENHAIN ND 287 User Manual User Manual
Page 121

ND 287
121
II – 7 Input and output of par
a
met
e
r list and er
ro
r
compensation table
P68
NOM.VALUE 0 =
+
55.0000
NOM.VALUE 1 =
+
0.0000
NOM.VALUE 2 =
+
0.0000
NOM.VALUE 3 =
+
0.0000
NOM.VALUE 4 =
+
1.0000
NOM.VALUE 5 =
+
0.0000
NOM.VALUE 6 =
+
0.0000
NOM.VALUE 7 =
+
0.0000
NOM.VALUE 8 =
+
0.1000
NOM.VALUE 9 =
+
13.9700
Nominal dimension for part 0 – part 9: 0.0000 is the default setting
P69
UP. LIMIT 0 =
+
60.0000
UP. LIMIT 1 =
+
0.0000
UP. LIMIT 2 =
+
0.0170
UP. LIMIT 3 =
+
25.0000
UP. LIMIT 4 =
+
5.0000
UP. LIMIT 5 =
+
6.0000
UP. LIMIT 6 =
+
7.0000
UP. LIMIT 7 =
+
7.0000
UP. LIMIT 8 =
+
9.0000
UP. LIMIT 9 =
+
254.0000
Upper tolerance limit for part 0 – part 9: 0.0000 is the default setting
P70
SORTING =
1
Sorting and tolerance checking: 0 = off, 1 = on
P71
POS-SPEED =
1
Measured value recording: 0 = position, 1 = speed
P73 SORT. COLOR =
1
Display color during sorting: 0 = blue , 1 = red, green
P74
EXT. INPUTS =
0
Function of external inputs: 0 = version 1, 1 = version 2 (X1+X2...)
P75 SERIES.MEAS. =
2
Measurement series display: 0 = off 1 = MIN, 2 = ACTL, 3 = MAX, 4 = DIFF
P76 RECORD VAL. =
1
Measured value recording: 0 = off, 1 = on
P77
LATCH =
2
Latching: 0 = interval, 2 = external signal, 3 = ENTER key
P78 NUMBER VAL. =
10
Number of measured values: 10 (0 - 10000), 0 is the default setting
P79
TIME SEC =
5
Time slot for measurement series in seconds: 5 s, 0 s is the default setting
P80
TIME MIN =
0
Time slot for measurements series in minutes: 0 min is the default setting
P81
TIME H =
0
Time slot for measurement series in hours: 0 h is the default setting
P82
INTERVALL =
0
Sampling interval for measurement series: 20 ms - 10 sec, 0 ms is the
default setting
P83
MEAS./SPC =
1
Measurement series / SPC: 1 = SPC active, 0 = measurement series active
P84
LATCH SPC =
0
Measured value latching (SPC): 0 = ENTER key, 1 = external signal
P85 MODEL SPC =
0
SPC distribution model: 0 = symmetrical, 1 = left-hand limit, 2 = right-hand
limit
P86 NR. SAMPLE =
25
Number of samples: 25 is the default setting
P87 VAL./SAMPLE =
3
Number of measured values per sample: 5 is the default setting
Parameter
Meaning