Examples of parameter lists, Nd 287 with rotary encoder connected to input x1 – HEIDENHAIN ND 287 User Manual User Manual
Page 118
![background image](https://www.manualsdir.com/files/815450/content/doc118.png)
118
II Installation, specifications
II – 7 Input and output of par
a
met
e
r list and er
ro
r
compensation table
Examples of parameter lists
ND 287 with rotary encoder connected to input X1
The parameter text is always sent in English.
The parameter value, not the text, is definitive when the parameters are transferred to the ND. The values in
boldface type are the default settings
.
Parameter
Meaning
#
Start character (#)
ND-287 1
DEG
Model: ND-287, Unit of measure: DEG, DMS or rad
P01
LINEAR =
0
Unit of measure for linear dimensions: mm = 0, inches = 1
P02
ANGULAR =
0
Unit of measure for angular dimensions: DEG = 0, DMS = 1, rad = 2
P03
ENC. TYPE =
1
Encoder type: Linear = 0, Rotary = 1
P04 ENC. SIGNAL =
1
Encoder signal: 0 = 11 µA, 1 = 1 Vpp, 2 = EnDat, 3 = ANALOG
P05
AXES DISPL. =
0
Display: 0 = X1, 1 = X2, 2 = X1 + X2, 3 = X1 - X2, 4 = f(X1,X2)
P06
ANGLE =
0
Angle format: 0 = +/– 180°, 1 = 360° , 2 = +/– infinity
P10
SCALING =
0
Scaling: 0 = off, 1 = on
P11 SCL. FACTOR = + 1.000000
Scale factor = 1.000000 (default setting)
P20 BRIGHTNESS =
94
Brightness of LCD screen: 0 - 100 % (80 % is the default setting)
P21
DISP. SAVER =
1
Screen saver: 0 = off, 1 = on
P22 SAVER TIME =
120
Idle time before screen saver is activated: 120 min (default setting)
P23 START.DISPL. =
1
Power-up screen: 0 = off, 1 = on
P24
SOFTKEY.X12 =
10101
Display mode selectable with SOFTKEY X1/X2: 11111 = all are activated
P30
DIRECTION =
0
Count direction: 0 = positive, 1 = negative
P31 SIGN.PERIOD =
20
Signal period: 20 µm (10 µm is the default setting)
P32
SP/R =
36000
Signal periods per revolution: 36000 is the default setting
P33 COUNT MODE =
5
Counting mode: 0 - 5 = 5, 0 - 2 = 2, 0 - 1 = 1
P34
DP PLACES =
4
Decimal places: 4 (default setting)
P35
REF ON/OFF =
1
Reference mark: 0 = off, 1 = on
P36
REF MARK =
5
0 = single reference mark
, 1..6: coded reference marks
P37
ALARM =
3
0 = off, 1 = frequency, 2 = contamination, 3 = frequency and
contamination
P38
EXT. REF =
1
External REF input: 0 = inactive, 1 = active