II Installation, specifications
II – 2 Installation Set
u
p
Encoder test
Use this test to check the signals of the 11 µApp or 1 Vpp interface,
the EnDat interface or the voltage at the analog module.
If you have configured the ND for two axes in the APPLICATION
field under COUNTER SETTINGS, you can perform the test for the
two possible encoder inputs.
Select input X1 or X2 and press ENTER.
Encoders with 11 µApp or 1 Vpp interface:
As the encoder is moved, you will see the signals of the A and B
channels as Lissajous figure (see Fig. II.49), as well as the
amplitude, symmetry and phase offset values.
Fig. II.49 Encoder test: 1 Vpp interface
Fig. II.50 Encoder test: 11 µApp interface