I Working with the ND 287 position display unit

I – 6 Sor

ting and t

o

ler

a

nce c

h

ec

king

I – 6 Sorting and tolerance checking

Sorting function

To sort the parts, the ND 287 compares the displayed value with an

upper or lower sorting limit, and displays the sorting result in color as

a display value and with the appropriate sorting symbol in the status

display:

Green

symbol: =

Values that are displayed in green are within the sorting limits.

Red

symbol: < or >

Values that are displayed in red are smaller than the defined lower

sorting limit or greater than the defined upper sorting limit.

At the same time, the sorting result is transmitted via two switching

outputs (pins 17 and 18) at connection X41 (see "Switching outputs at

D-sub connection X41" on page 95).

Pin 17

becomes active if the displayed value is smaller than the

lower sorting limit.

Pin 18

becomes active if the displayed value is greater than the

upper sorting limit.

The sorting and tolerance checking mode sorts the display value into

one of the three following classes:

Within tolerance

Above tolerance

Below tolerance

This means that the sorting limits correspond to the tolerance limits of

SPC.

Fig. I.66 Status display for sorting and tolerance

checking