Setting up the serial port, Setting up the data interface, Ii – 2 installation set u p – HEIDENHAIN ND 287 User Manual User Manual
Page 85

ND 287
85
II – 2 Installation Set
u
p
Setting up the serial port
The ND 287 features two serial interfaces: RS-232-C/V.24 (X31) and
USB (UART, X32).
A printer or computer with a serial data interface may be connected to
the available serial ports in order to perform the following tasks:
Sending measured values, error compensation tables or
configuration files to a printer or computer.
Receiving error compensation tables or configuration files from a
computer.
In addition, the interfaces allow external operation of the ND 287.
As an option, an Ethernet module (100baseT) can be connected to
X26/X27 so that an Ethernet interface is available.
Setting up the data interface
In the INSTALLATION SETUP menu, select INTERFACE SETTINGS.
In the SERIAL PORT field, use the USB/RS-232 soft key to select the
interface you want to use.
The BAUD RATE field can be set to 110, 150, 300, 600, 1200, 2400,
9600, 19200, 38400, 57600 or 115200 by using the DECREASE or
INCREASE soft keys.
Fig. II.41 Connections
X1
X32/X31
Danger of electrical shock!
The interfaces X31 and X32 comply with the requirements
of EN 50 178 for low voltage electrical separation.
Do not engage or disengage any connecting elements
while the unit is under power!
Fig. II.42 RS-232-C/V.24 data interface
Fig. II.43 USB (UART) data interface