ND 287

53

I – 5 Ser

ies of measur

ements and statistical

pr

ocess

c

ontr

ol

Tolerances

In the TOLERANCES form, you define the tolerance limits for

statistical process control:

In the SPC SETUP submenu, select TOLERANCES and press the

ENTER key to open the form.

The currently selected part (P0-P9) is displayed at the upper right.

You can also change the tolerances for the selected part in the

SELECT PART form that is opened by pressing the PART

TOLERANCE soft key on the standard screen.

Use the numeric keys to enter the values for the lower tolerance
limit, the nominal value and the upper tolerance limit

in the

LOWER LIMIT, NOMINAL VALUE and UPPER LIMIT fields. The

values for the lower and upper tolerance limits correspond to the

values for the lower and upper sorting limits. This means that the

ND activates pins 17 or 18 at D-sub connection X41 if the limits are

violated (see "Sorting and tolerance checking" on page 58).

Press the RED, GREEN/[DISPL.BLUE] soft key to select whether

the display value changes to the same color as the sorting symbols.

Blue is the default setting for the color of the display value (see

"Sorting and tolerance checking" on page 58).

If you edit the values entered, the ND displays a warning.
In order to save the changes, the ND must clear the
data stored in the FIFO memory.

Press the ENTER key

to confirm, or press the C key to cancel.

Fig. I.58 Tolerances

Please note that the LOWER LIMIT must be less than the

NOMINAL VALUE and the UPPER LIMIT, and the UPPER

LIMIT must be greater than the NOMINAL VALUE.