beautypg.com

Defining the display for a series of measurements – HEIDENHAIN ND 287 User Manual User Manual

Page 46

background image

46

I Working with the ND 287 position display unit

I – 5 Ser

ies of measur

ements and statistical

pr

ocess

c

ontr

ol

Defining the display for a series of
measurements

In the SERIES OF MEASUREMENTS menu, select MEAS. SERIES

DISPLAY. Then press the DISPLAY MEAS. SER. soft key to select the

value to be displayed while a series of measurements is being

performed:

ACTL DISPLAY: Current measured value is displayed.

MIN DISPLAY: Minimum value of a series of measurements is

displayed.

MAX DISPLAY: Maximum value of a series of measurements is

displayed.

DIFF DISPLAY: Difference MAX - MIN (i. e. the range) is displayed.

Fig. I.42 Display for series of measurements

Fig. I.43 MIN, MAX and DIFF of a curved surface

Fig. I.44 Series of measurements for determining

eccentricity

DIFF

MAX

MIN

ACTL

2e = r

MAX

– r

MIN

e = DIFF

2

r

MAX

r

MIN

e