I Working with the ND 287 position display unit
I – 5 Ser
ies of measur
ements and statistical
pr
ocess
c
ontr
ol
Defining the display for a series of
measurements
In the SERIES OF MEASUREMENTS menu, select MEAS. SERIES
DISPLAY. Then press the DISPLAY MEAS. SER. soft key to select the
value to be displayed while a series of measurements is being
performed:
ACTL DISPLAY: Current measured value is displayed.
MIN DISPLAY: Minimum value of a series of measurements is
displayed.
MAX DISPLAY: Maximum value of a series of measurements is
displayed.
DIFF DISPLAY: Difference MAX - MIN (i. e. the range) is displayed.
Fig. I.42 Display for series of measurements
Fig. I.43 MIN, MAX and DIFF of a curved surface
Fig. I.44 Series of measurements for determining
eccentricity
DIFF
MAX
MIN
ACTL
2e = r
MAX
r
MIN
e = DIFF
2
r
MAX
r
MIN
e