II Installation, specifications

II – 7 Input and output of par

a

met

e

r list and er

ro

r

compensation table

P34.1 DP PLACES =

4

X1: Decimal places: 4 is the default setting

P34.2 DP PLACES =

4

X2: Decimal places: 4 is the default setting

P35.1 REF ON/OFF =

1

X1: Reference mark: 0 = off, 1 = on

P35.2 REF ON/OFF =

1

X2: Reference mark: 0 = off, 1 = on

P36.1 REF MARK =

5

X1: 0 = single reference mark, 1..6: coded reference marks

P36.2 REF MARK =

5

X2: 0 = single reference mark, 1..6: coded reference marks

P37.1

ALARM =

3

X1: 0 = off, 1 = frequency, 2 = contamination, 3 = frequency and
contamination

P37.2

ALARM =

3

X2: 0 = off, 1 = frequency, 2 = contamination, 3 = frequency and
contamination

P38

EXT. REF =

1

External REF input: 0 = inactive, 1 = active

P39.1 SCREW.PITCH =

88.123456

X1: Screw pitch in mm (multiturn rotary enc.), 10 mm is the default setting

P39.2 SCREW.PITCH =

10

X2: Screw pitch in mm (multiturn rotary enc.), 10 mm is the default setting

P40.1 ENC. COMP. =

2

X1: Axis error compensation: 0 = off, 1 = linear, 2 = non-linear

P40.2 ENC. COMP. =

2

X2: Axis error compensation: 0 = off, 1 = linear, 2 = non-linear

P41.1 LIN. COMP. =

+

0.0

X1: Linear compensation: 0.0 µm/m (default setting)

P41.2 LIN. COMP. =

+

0.0

X2: Linear compensation: 0.0 µm/m (default setting)

P43 .1 ANALOG U1 = +

10.000

X1: Analog module: Voltage 1 = 10.000 V (default setting)

P43 .2 ANALOG U1 = +

10.000

X2: Analog module: Voltage 1 = 10.000 V (default setting)

P44.1 ANALOG U2 = -

10.000

X1: Analog module: Voltage 2 = –10.000 V (default setting)

P44.2 ANALOG U2 = -

10.000

X2: Analog module: Voltage 2 = –10.000 V (default setting)

P45.1 ANALOG.POS1 = +

10.0000

X1: Analog module: Position 1 (10.000 is the default setting)

P45.2 ANALOG.POS1 = +

10.0000

X2: Analog module: Position 1 (10.000 is the default setting)

P46.1 ANALOG.POS2 = -

10.0000

X1: Analog module: Position 2 (–10.000 is the default setting)

P46.2 ANALOG.POS2 = -

10.0000

X2: Analog module: Position 2 (–10.000 is the default setting)

P47 ANALOG FCT = +

9.4

Temperature compensation: Coefficient +9.4 µ/K (default setting)

P48

REF. TEMP. = +

20.00

Temperature compensation: Reference temperature +20.0 ° (default

setting)

P49 ANALOG.COMP. =

1

Temperature compensation: 0 = off, 1 = on

P50

RS232/USB =

1

Interface: 0 = RS232, 1= USB

P51 BAUD RATE =

11

Baud rate = 115200 (0 - 11), 7 is the default setting

Parameter

Meaning