II – 2 Installation Set

u

p

Incremental rotary encoder

For rotary encoders, enter the encoder's signal period per revolution

(360°) directly in the SIGNAL PERIOD field (see "Encoder

parameters" on page 98). Press the DOWN arrow key to move to the

next parameter.

In the REFERENCE MARK field, use the numeric keys to enter the

number of reference marks (0 for NONE, 1 for SINGLE, ...) per

revolution (360°).

In the EXTERNAL REF field, press the ON/OFF soft key to select

whether pin 25 at connection X41 is active or inactive. This pin can

be used to activate or deactivate the reference mode. The

respective current status is changed.

In the COUNT DIRECTION field, select the count direction by

pressing the POSITIVE/ NEGATIVE soft key. If the encoder's count

direction matches the user's traverse direction, select POSITIVE. If

the directions do not match, select NEGATIVE.

In the ERROR MONITOR field, select whether the ND will monitor

and display encoder counting and signal errors by pressing the

ERROR soft key. In the ERROR MONITOR field, you can select OFF,

FREQUENCY, CONTAMINATION or FRQ.+CONTAMINATION.

When an error message occurs, press the C key to acknowledge it.

Counting errors

are caused by contamination or excessive

frequency:

Contamination

: The signal to the encoder falls below a set limit.

Frequency errors

: The signal frequency exceeds the set limit.

Fig. II.13 Incremental rotary encoder form

Fig. II.14 Incremental rotary encoder form

Fig. II.15 Incremental rotary encoder form

If you set the ERROR MONITOR field to OFF, the ND 287

will ignore encoder errors.