I Working with the ND 287 position display unit

I – 5 Ser

ies of measur

ements and statistical

pr

ocess

c

ontr

ol

Setting up a series of measurements

To define the parameters of the series of measurements:

Call the SERIES OF MEASUREMENTS menu.

Select MEAS. SERIES SETUP.

In the RECORD MEAS. VALUES field, activate or deactivate

measured-value recording for a series of measurements.

In the RECORD field, select the trigger for recording the measured

values of a series of measurements. Press the RECORD soft key to

select one of the following possibilities:

Sampling interval

External signal at connection X41 (pin 22 or 23)

ENTER key

Press the DOWN arrow key or the NAVIGATION key to display

further parameters.

If you have selected EXTERNAL SIGNAL or ENTER key, another

parameter needs to be defined:

In the NO. MEAS. VALUES field, you enter the number of measured

values for the series of measurements. If you enter the value zero,

the ND deactivates the RECORD MEASURED VALUES function.

Fig. I.39 Setting up a series of measurements

Fig. I.40 Setting up a series of measurements

The ND 287 can store a maximum of 10000 measured

values per axis. The measured values recorded in a series

of measurements will only be retained in memory until

power is switched off.