Duration of measured value transfer, Ii – 6 measur ed v alue output – HEIDENHAIN ND 287 User Manual User Manual

Page 115

background image

ND 287

115

II – 6 Measur

ed v

alue output

Duration of measured value transfer

Example: Data sequence during measured-value output
Measured value: X = - 5.23 mm
The measured value is within the sorting limits (=) and is the current

value (A) of a series of measurements.
Measured value output:

t

D

Duration of measured value transfer in [s]

L: Number of blank lines
B: Baud rate

-

5.23

=

A

1

2

3

4

5

6

7

8

1

+/– sign

2

Numerical value with decimal point: 10 characters on the whole,

leading zeros are output as blank spaces.

3

Blank space

4

Unit of measure: Blank space = mm, = inches, G = degrees,
M

= DMS, R = rad, m = mm/min, i = ipm, U = rpm, ? = fault

5

Sorting status (< / = / >)
?

= lower sorting limit > upper sorting limit

6

When series of measurements has been started:
S

= MIN, A = ACTL, G = MAX, D = DIFF

If two axes are coupled (optional) and no series of

measurements has been started:
1

= X1, 2 = X2, A = X1 + X2, S = X1 - X2, F = f(X1,X2)

7

Carriage return

8

Blank line (Line Feed)

t

D

187

11 L

(

)

+

B

----------------------------

=