Duration of measured value transfer, Ii – 6 measur ed v alue output – HEIDENHAIN ND 287 User Manual User Manual
Page 115
ND 287
115
II – 6 Measur
ed v
alue output
Duration of measured value transfer
Example: Data sequence during measured-value output
Measured value: X = - 5.23 mm
The measured value is within the sorting limits (=) and is the current
value (A) of a series of measurements.
Measured value output:
t
D
Duration of measured value transfer in [s]
L: Number of blank lines
B: Baud rate
-
5.23
=
A
1
2
3
4
5
6
7
8
1
+/– sign
2
Numerical value with decimal point: 10 characters on the whole,
leading zeros are output as blank spaces.
3
Blank space
4
Unit of measure: Blank space = mm, “= inches, G = degrees,
M
= DMS, R = rad, m = mm/min, i = ipm, U = rpm, ? = fault
5
Sorting status (< / = / >)
?
= lower sorting limit > upper sorting limit
6
When series of measurements has been started:
S
= MIN, A = ACTL, G = MAX, D = DIFF
If two axes are coupled (optional) and no series of
measurements has been started:
1
= X1, 2 = X2, A = X1 + X2, S = X1 - X2, F = f(X1,X2)
7
Carriage return
8
Blank line (Line Feed)
t
D
187
11 L
•
(
)
+
B
----------------------------
=