Measured-value output, I – 4 j ob set up – HEIDENHAIN ND 287 User Manual User Manual

Page 39

background image

ND 287

39

I – 4 J

ob Set

up

Measured-value output

With the measured value output feature, the current display values

can be transmitted over the serial port. Output of the current display

values is activated via a switching signal at D-sub connection X41,

the command CTRL B, or the PRINT soft key (see "Measured value

output" on page 113).
The effect the signal for measured value output has on the display

value on the screen can be defined as follows:

In the JOB SETUP menu, select MEASURED VALUE OUTPUT and

press the ENTER key to open the form.

Press the DISPLAY FREEZE soft key. Three different settings are

available:

CONCURRENT DISPLAY: The display is not stopped during

measured value output. The display value is the current measured

value.

FROZEN/CONCURRENT DISPLAY: The display is stopped during

measured value output and remains stopped while the switching

input is active.

FROZEN DISPLAY: The display is stopped, but is updated by

every measured value output.

The following figure illustrates the different settings for measured

value output.

For more information on measured value output, refer to Page 113.

Fig. I.32 Measured-value output

1

2

3

4

1

Position

2

Latch signal

3

Frozen display

4

Frozen/concurrent display