3.9 Performing Calibration Correction (/CC1
Option)
The input value is calibrated using segments, and the result is used as a measured
value.
* You cannot specify calibration correction for a channel whose log scale (/LG1 option) mode
is set to LogType2.
For a description of the function, see section 1.1.
Setup Screen
Press
MENU (to switch to setting mode), and select the Menu tab > Meas channel >
Calibration correction.
Setup Items
True val
Number of set points
Calibration set points
Y5
Y4
Y3
Y2
Y1
MES val
X1 X2 X3 X4
X5
Condition
X1 < X2 < X3 < X4 < X5
• First-CH, Last-CH
Select the target channels. You can set consecutive channels whose range is set to
the same value as the first channel.
• Number of set points
Select the number of points that make up the segments (including the start and end
points) in the range of 2 to 16.
To disable calibration correction, select
Off.
• Number of set points > MES val, True val
Press the
Input soft key and enter the value.
For the MES value, set a value that is greater than the previous value.
Press the
Measure soft key to set the measured value at that point to MES val. If
you press the
Measure soft key when setting multiple channels simultaneously, the
measured value of the first channel are set to the
MES val of all channels.
Selectable Range of MES and True Values
• Channels on which linear scaling is specified
–30000 to 30000 (the decimal place is the same setting as the scale value)
• Other channels
Value in the measurable range of the selected range
Example: –2.0000 to 2.0000 for 2 V range
Note
• The calibration correction setting is set to Off if you change the
Mode or Range setting.
• Calibration correction cannot be specified on channels set to
Skip.