10-18

IM 04L21B01-01EN

Does the input meet

the specifications?

NO

NO

NO

NO

NO

YES

YES

YES

YES

• Large measurement errors
• Trend or digital wave value fluctuation
• Trend going off the scale beyond either 0% or 100%

YES

• Adjust the span and range

settings (section 3.3).

• If you use input calibration

(/CC1 option), check the set values

for correction points (section 3.9).

Change the input wiring to meet

the specifications.

Is the input wiring correct?

• Attach the input terminal cover

correctly.

• Protect the input terminal from fans

and other sources of wind.

• Keep the temperature fluctuation

near the input terminal to a

minimum.

Are the range/span

settings appropriate?

Are you taking

measures against noise?

Are

you keeping

the environmental temperature

constant (TC

input)?

• Separate the input wiring from the

noise source.

• Ground the recorder.

• Ground the measured object.

• Insulate the TC from the measured

object.

• Use shielded cables for input.

• Change the integration time

of the input A/D converter

(section 3.1).

• Use the moving average of the

input (section 3.4).

• Wire correctly

• Install the terminal block correctly.

• Fasten the terminal screws

correctly.

• Insulate the RTD from the ground.

• If burnt, replace the TC.

Is the input wiring parallel

with other equipment?

NO

NO

YES

YES

Is the RJC setting

correct (TC input)?

• Stop the TC burnout upscale/

downscale function of other

equipment.

• Ground the FX and other

equipment at one point.

• Disconnect parallel wiring

(for example, use a

double-element TC)

Configure correctly.

Contact your nearest YOKOGAWA
dealer to have your FX serviced.

Section 3.2

Section 13.1

10.2 Troubleshooting