App index action – Yokogawa Value Series FX1000 User Manual
Page 44
1-35
IM 04L21B01-01EN
Overview of Functions
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
Action
• Actions
Select from the following actions.
Action
Level/Edge
1
Description
Memory start/stop
Level
Starts/stops memory sampling.
Memory start
Edge
Starts the memory sampling.
Memory stop
Edge
Stops the memory sampling.
Event trigger
Edge
Applies a trigger for starting the event data
recording.
This is valid when recording event data in trigger
mode. See the next page.
Alarm acknowledge
Edge
Releases the alarm output.
This is valid when the use of the alarm ACK
operation is enabled.
Computation start/stop
2
Level
It is used to start/stop the computation.
Computation start
2
Edge
Starts the computation.
Computation stop
2
Edge
Stops the computation.
Computation reset
2
Edge
Resets the computed values on all computation
channels.
Save display data
Edge
The display data being recorded is saved to the
internal memory as a file. This is the same function
as the data save operation using the FUNC key.
Save event data
Edge
The event data being recorded is saved to the
internal memory as a file.
This is the same function as the data save operation
using the FUNC key.
Message
Edge
Writes a message.
This action can be executed while memory sampling
is in progress.
Snapshot
Edge
Saves the screen image data.
Switch the display rate
Level
Toggles between the trend interval and the
secondary trend interval.
This action is valid when the FX is configured to use
trend interval switching.
Manual sample
Edge
Executes manual sampling.
Reset the relative timer
Edge
Resets the relative timer. The timer restarts from
that point. See the next page.
Switch the display group
Edge
Switches the display group when the trend, digital,
or bar graph is displayed.
Flag
2
Level
The flag is zero for normal conditions and 1 when an
event occurs. When the event is an edge operation,
the value alternates between 0 and 1 whenever an
event occurs.
The flag can be written in a calculation expression
of a computation channel.
Load the setup
Edge
Loads the setup data file in the root directory of the
CF card into the FX and updates the FX settings.
See the next page.
Adjust the time
Edge
Synchronizes the time to the nearest hour. See the
next page.
1 For a description of level and edge, see “Miscellaneous” in this section.
2 This is an option.
1.6 Event Action and Remote Control Functions (/R1 and /PM1 Options)