6.4 Starting and Stopping Recording and Saving

Measured Data

This section explains how to start recording and save measured data to the internal
memory.
You can also save the measured data that is stored in internal memory to an external
storage medium.

* This is only valid on FXs that have a CF card slot or USB interface (/USB1 option).

For a description of the function, see section 1.4.

Procedure

• Starting the Recording (Memory Start)

Press

START. The internal memory icon in the status display section changes from

the stop icon to memory sampling icon.
• When recording display data or event data in free mode, recording starts.
• When recording event data in trigger mode, the FX enters the trigger-wait condition.

• Applying a Trigger to Start the Recording

Carry out the procedure below when the FX is waiting for a trigger.

Trigger through Key Operation

The procedure below can be carried out when recording event data in trigger mode

and the FX is configured so that the start trigger is applied through key operation.

1.

Press

FUNC.

The Function menu appears.

2.

Press the

Trigger soft key.

The recording starts.

Trigger by an Event (Event action function must be configured. See

section 7.1.)

Recording starts when an event occurs.

• Automatically Saving Measured Data

Automatic saving takes place when

Auto save is set to On (see section 6.2 for details).

The save destination is the CF card.

* This is only valid on FXs that have a CF card slot.

Have the CF card inserted in the slot at all times. While the memory sampling is in

progress, the measured data recorded in the internal memory is automatically saved
to the CF card.

Action when Media FIFO is not enabled: If data is not completely saved to the external

storage medium such as due to insufficient free space, the next time that data is
automatically saved, the unsaved data will be saved.