3.8 Releasing the Alarm Output (Alarm ACK

Operation)

This operation is valid when the FX is set as follows:
• When

Indicator is set to Hold in the basic setting mode.

• When

Relay Hold is set to Hold and Relay action on ACK is set to Normal in the

basic setting mode.

• When

Relay action on ACK is set to Reset in the basic setting mode.

For the Hold and Relay Action on ACK setting procedures, see section 3.5.

Procedure

This operation is carried out after an alarm occurs.

1.

In Operation Mode, press

FUNC.

The Function menu appears.

2.

Press the

AlarmACK soft key.

The alarm output is released.

Explanation

• Alarm Acknowledge (ACK) Operation

When an alarm acknowledge operation is carried out, the indicators and outputs (relays

and switches) of all activated alarms are cleared.