Troubleshooting

3

2

1

4

5

6

7

8

9

10

11

12

13

14

App

Index

Taking measures
against noise?

NO

YES

• Move the input and communication wires
away from the noise source.
• Ground the recorder.
• Use shielded cables for input and
communication wires.
• Insert an isolation transformer into the
power line.
• Insert a power line filter into the power line.
• If an inductive load is used in the alarm
output line, insert a serge suppressor in
the line.

Contact your nearest YOKOGAWA
dealer.

Display and other functions do not work properly.

Are all of

the typical conditions for

displaying the item

satisfied?

NO

YES

• Make sure that the menu item is set
to be displayed using the menu
customize function.
If the item is not set to be displayed,
then set it to be displayed
(see section 5.16).
• Confirm that the FX is equipped
with the options required for that
menu item.

Contact your nearest YOKOGAWA
dealer to have your FX serviced.

Items on the display selection menu or

function menu are not displayed.

10.2 Troubleshooting