5.9 Partially Expanding the Waveform

Expand a part of a waveform (shrink the other parts) on the display.

* You cannot specify a channel that is set to Log scale (/LG1 option).

We recommend that you display the scale when viewing partially expanded channels.
The numbers for the ends of the scale and the boundary position are displayed, and
you can identify the expanded and reduced areas easily. However, numbers are not
displayed for other scale marks.
For a description of the function, see section 1.3.

Setup Screen

• Turning ON/OFF the Partial Expanded Display Function

Press

MENU (to switch to setting mode), hold down FUNC for 3 s (to switch to basic

setting mode), and select the

Environment tab > View, Message.

• Partially Expanded Display Method

Press

MENU (to switch to setting mode), and select the Menu tab > Meas channel >

Partial.

Setup Items

• View > Partial

If you select

On, the Partial setup item appears in the setting mode.

• First-CH, Last-CH

Select the target channels.

• Partial > On/Off

To enable partial expanded display, select

On.

• Partial > Expand

Set the position where the value specified by

Boundary is to be displayed within the

display span in the range of 1 to 99.