Chapter 13 specifications, 1 signal input and alarms, 1 signal input and alarms -1 – Yokogawa Value Series FX1000 User Manual
Page 266: App index, Measurement input
13-1
IM 04L21B01-01EN
Specifications
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
13.1 Signal Input and Alarms
Measurement Input
Item
Specifications
Number of inputs, scan interval, and A/D integration time
Model
No. of
Measurement Channels
Scan Interval
A/D converter integration
time
FX1002
2
125 ms, 250 ms
AUTO
1
, 50 Hz, 60 Hz
FX1004
4
FX1006
6
1 s, 2 s, 5 s
AUTO
1
, 50 Hz, 60 Hz,
100 ms
2
FX1008
8
FX1010
10
FX1012
12
1 AUTO: The FX automatically switches between 50 Hz and 60 Hz depending on the power
supply frequency.
2 You can only set the integration time to 100 ms on models FX1006 through FX1012. If you
set the integration time to 100 ms, you can only set the scan interval to 2 or 5 seconds.
Input Type
DC voltage, 1-5V, thermocouple (TC), resistance temperature detector (RTD), ON/OFF input (DI),
and DC current (by adding an external shut resistor)
Measurement range and measurable range
Input Type
Range
Measurable Range
DC voltage
20 mV
–20.000 to 20.000 mV
60 mV
–60.00 to 60.00 mV
200 mV
–200.00 to 200.00 mV
1 V
–1.0000 to 1.0000 V
2 V
–2.0000 to 2.0000 V
6 V
–6.000 to 6.000 V
20 V
–20.000 to 20.000 V
50 V
–50.00 to 50.00 V
1-5V
1 to 5 V
6
0.800 to 5.200 V
Thermocouple
R
1
0.0 to 1760.0°C
32.0 to 3200.0°F
S
1
0.0 to 1760.0°C
32.0 to 3200.0°F
B
1
0.0 to 1820.0°C
32.0 to 3308.0°F
K
1
–200.0 to 1370.0°C
–328.0 to 2498.0°F
E
1
–200.0 to 800.0°C
–328.0 to 1472.0°F
J
1
–200.0 to 1100.0°C
–328.0 to 2012.0°F
T
1
–200.0 to 400.0°C
–328.0 to 752.0°F
N
1
–270.0 to 1300.0°C
–454.0 to 2372.0°F
W
2
0.0 to 2315.0°C
32.0 to 4199.0°F
L
3
–200.0 to 900.0°C
–328.0 to 1652.0°F
U
3
–200.0 to 400.0°C
–328.0 to 752.0°F
WRe
4
0.0 to 2400.0°C
32.0 to 4352.0°F
RTD
Pt (Pt100)
5
–200.0 to 600.0°C
–328.0 to 1112.0°F
JPt(JPt100)
5
–200.0 to 550.0°C
–328.0 to 1022.0°F
DI
Level
0: Less than 2.4 V. 1: 2.4 V or higher (judged at the 6 V range)
Contact
7
0: Open. 1: Closed (parallel capacitance of 0.01 μF or less)
1 R, S, B, K, E, J, T, N: IEC584-1 (1995), DIN IEC584, JIS C1602-1995
2 W: W-5%Re/W-26%Re (Hoskins Mfg. Co.), ASTM E988
3 L: Fe–CuNi, DIN43710, U: Cu–CuNi, DIN43710
4 WRe: W-3%Re/W-25%Re (Hoskins Mfg. Co.)
5 Pt100: JIS C1604-1997, IEC751-1995, DIN IEC751-1996
JPt100: JIS C1604-1989, JIS C1606-1989. Measuring current i = 1 mA (Pt100, JPt100).
6 The range for linear scaling of 1-5V inputs. Burnout detection and low-cut functions are
available.
7 The detected current value is approximately 10 μA.
Chapter 13 Specifications