3.13 Using the Log Scale to Perform Measurements (/LG1 Option)

Note

• Set the mantissa to 1.00.
• The output voltage for pressure reading whose mantissa is 1.00 has a certain deviation.
For example, if the pressure reading is 1.00E-5, the output voltage is 2.00 to 2.10 V. In this

situation, set the range to the minimum value (2.00 V).

• Setting the Nonlinear Log Input (LogType1)

To support the nonlinear output of vacuum gauges, use the calibration correction

function (/CC1) to correct the recorder input values with linear approximation.

Conceptual Diagram

The pressure scale is mapped to an input scale that has been multiplied by a constant

number to derive the output scale.

10

9

8

7

6

5

4

3

2

1

0

1

2

3

4

5

6

7

8

9

10

10

-1

10

0

10

1

0

10

2

10

3

Input value [V]

Pressure scale

Output scale

Output value [V]

1

Press

MENU (to switch to setting mode), and select the Menu tab > Meas channel

>

Range, Alarm.

Set the range as shown below. To use the calibration correction function (/CC1

option), be sure to set the mode to LogType1.

Mode: LogType1
Range: 20V
Span: Lower 0.000, Upper 10.000
Scale: Lower 1.00E-1, Upper 4.00E+3

For the span and scale, assign the boundary values of the output scale and pressure

scale in the conceptual diagram.