• Report channel number
The report is output in order by this number.
• Report channel number > On/Off
Select
On for the report channels to be used.
• Report channel number > Channel
Set the channel to assign to the report channel. All channels can be assigned, but
reports are not created for channels set to
Skip or Off even if they are assigned.
Errors are returned for any reports for channels that are set to Log scale (/LG1 option).
In the stacked bar graph display (see section 4.10 for details), report data is displayed
in the following groups. However, only channels that have the same unit as the first
channel in the group are displayed.
Report Group
FX1002 and FX1004 FX1006, FX1008, FX1010, and FX1012
1
R01 to R06
R01 to R06
2
R07 to R12
R07 to R12
3
-
R13 to R18
4
-
R19 to R24
• Report channel number > Sum scale
Set the sum scale to
/s to /day to match the unit of the measured value.
Example: If the unit of the measured value is “m
3
/min,” select
/min.
Off: Sums as-is the measured data per scan interval.
• Handling of Overflow Data
Overflow data is handled in the same way as it is in statistical computations (TLOG
and CLOG).
See section 9.1.
Procedure
• Starting/Stopping the Report Function
Starting the memory sampling starts the report function. Likewise, stopping the
memory sampling stops the report function.
• Displaying the Reports
See section 4.5.
• Saving the Reports
See section 1.4.
9.5 Creating Reports