Yokogawa Value Series FX1000 User Manual

Page 281

background image

13-16

IM 04L21B01-01EN

Item

Specifications

Measurement accuracy

The performance values listed here were recorded under the following standard operating

conditions: 23 ± 2°C, 55 ± 10%RH; power supply frequency: 50/60 Hz ± 1% or less; rated input: ±

1% or less, power factor: 1 ± 1% or less; warm-up time: 30 minutes or more; and a location in which

vibrations and other factors do not affect the operation of the instrument.

Active power (W): ±1.0% of range

Voltage (V):

±1.0% of range

Current (A):

±1.0% of range

Apparent power, reactive power, and power factor: Value calculated from the measured values ± 1

digit

Expressions (V and A are rms values)

Apparent Power (VA)

Reactive Power (Q)

(Without using the

reactive power

measurement method)

Power Factor(PF)

Single-

phase two-

wire system

VA = V × A

Q = ((VA)

2

– P

2

)

ΣP/ΣVA
(Without using the reactive

power measurement

method)

Single-

phase,

three-wire

system

VAi = Vi × Ai
i = 1, 2
ΣVA = VA1 + VA2

Qi = ((VAi)

2

– Pi

2

)

i = 1, 2
ΣQ = Q1 + Q2

Three-phase

three-wire

system

VAi = Vi × Ai
i = 1, 3

ΣVA = 3/2(VA1 + VA3)

Qi = ((VAi)

2

– Pi

2

)

i= 1, 3
ΣQ = Q1 + Q3

*

The FX’s apparent power (VA), reactive power (Q), power factor (PF), and phase (deg)

are determined from the voltage(V), current(A), and active power(P) by means of digital

computations. Therefore, for distorted signal input, the value obtained on the FX may

differ from that obtained on other instruments that use a different method.

*

Make sure that the voltage input is at least 10% of the rated value and the current input

is at least 5% of the rated value.

*

In the Σ Q computation, each phase’s Q value is computed as negative (–) if the current

input is leading the voltage input and as positive (+) if the current input is lagging the

voltage input.

Frequency: ±1.0 Hz

*

The frequency of the voltage line input to voltage P1 is output in units of Hz.

Response time

2 seconds

Continuous overload

Within the degree of accuracy (rated voltage and rated current × 1.2 applied for 2 hours)

Instantaneous overload

Within the degree of accuracy (rated voltage × 1.5 applied for 10 seconds, rated current × 2 applied

for 10 seconds, rated current × 10 applied for 3 seconds)

Dielectric strength

2500 VAC (50/60 Hz) for 1 minute (between the current input, voltage input, and earth)

Insulation resistance

100 MΩ (500 VDC between the current input, voltage input, and earth)

Computation functions

The same as the computation function (/M1)

Effects of the operating conditions
Ambient temperature

±0.05%/°C (under the following conditions: 0 to 50°C, 0.05 In ≤ I ≤ Imax, power factor = 1)

±0.07%/°C (under the following conditions: 0 to 50°C, 0.1 In ≤ I ≤ Imax, power factor = 0.5)

In: Rated current

Voltage variation

Within the degree of accuracy (90 to 132 VAC or 180 to 250 VAC; frequency is 50 or 60 Hz)

External magnetic fields 400 A/m or less

Active power and voltage: ±1.0% of range

Effect of the input frequency

For a change within 45 to 65 Hz, the effect on the active power, voltage, and current is within the

accuracies.

13.5 Options